Compact NDIR Gas Analyzer System
Simultaneous and continuous measurement of up to 5 components among: NOx, SO2, CO, CO2, CH4, and O2
Suited for monitoring of flue gas, combustion gas, biogas, etc.
- Continuous measurement for 5 days max.
- Separatable structure
- A variety of functions: auto calibration, O2-corrected output, averaging, alarm, etc.
- Easy-to-see LCD
You can check the concentration changes of all components at a glance - Easy-to-use interface
- You can select a paramagnetic O2 sensor or a galvanic O2 sensor
Category: Uncategorized
Description
Specifications
| Principle | NOX, SO2, CO, CO2, CH4: Non dispersive infrared sensor O2: paramagnetic sensor or galvanic sensor |
|
|---|---|---|
| Components | 1 component: NOX, SO2, CO, CO2, CH4 2 components: NOX+SO2, CO2 + CO, CH4 + CO, CO2 + CH4, NOX + CO 3 components: NOX + SO2 + CO, NOX + SO2 + CO2, NOX + CO2 + CO, SO2 + CO2 + CO, CH4 + CO2 + CO 4 components: NOX + SO2 + CO + CO2 O2 measurement can be added to above all. |
|
| Range | NOX: 0–500 ppm…5000 ppm SO2: 0–500 ppm…1% CO: 0–200 ppm…100% CO2: 0–200 ppm…100% CH4: 0–1000 ppm…100% O2: 0–5%…25% |
|
| No. of ranges | 3 (Maximum range ratio 1 : 5) | |
| Repeatability | ±0.5% FS | |
| Zero drift | ≤ ±1.0% FS per day | |
| Span drift | ≤ ±1.0% FS per day | |
| Linearity | ±2.0% FS | |
| Response time | ≤ 50 s (for 90% response) O2 measurement with galvanic sensor and SO2 measurement: ≤ 3 min |
|
| Output signal | 4–20 mA DC (linear) or 0–1 V DC (linear), up to 8 points, non-insulation (instantaneous value, O2-corrected instantaneous value, O2-corrected average value) |
|
| Display | backlit LCD, English or Japanese | |
| Functions | Auto calibration, O2 correction, averaging, output signal hold, automatic display off | |
| Communication | RS-232C (Modbus protocol) | |
| Operating environment | -5°C to +40°C, RH 90% or less | |
| Power supply voltage | 100–115 V AC or 200–240 V AC, 50/60 Hz | |
| Power consumption | Analyzer unit: approx. 150 VA Sampling unit: approx. 250 VA |
|
| Dimensions and weight | Analyzer unit: 365 (W) x 211 (H) x 514 (D) mm, approx. 12 kg Sampling unit: 365 (W) x 377 (H) x 514 (D) mm, approx. 18 kg |
|
| Gas conditions | Temperature | 0–50°C at inlet |
| Flow rate | 0.5 L/min ±0.2 L/min per optical system | |
| Dust | ≤ 50 mg/Nm3 | |
| Pressure | ≤ 3 kPa (gas outlet should be open to the atmospheric air) | |
| Mist | Unallowable | |
| Corrosive gas | HCl: ≤ 10 ppm | |
Note: this product is non-explosion-proof. The utmost caution must be exercised when handling explosive gas or hazardous gas.